CMA Operating Manual: Canon Mask Aligner
Introduction
Safety Highlights | Pre-operational Checklist | Before you enter the cleanroom | First setup after entering the cleanroom | Setup after the mercury lamp has warmed up |
Operating Instructions
Subroutine 1: First Mask Exposure | Subroutine 2: Manual Wafer Load | Log Off | Troubleshooting |
Figures
Figure 1: Right View | Figure 2: Top View | Figure 3: Left View | Figure 4: Main Panel | Figure 5: Front Panel | Figure 6: Mercury Lamp Power Supply | Figure 7: Wafer Carrier |
Introduction
Use the Canon PLA-501F Mask Aligner (See Fig. 1: Right View) whenever you have a process flow that requires more than one mask on a wafer. If it is critical to align these masks with each other, you must use this tool. The Northern California Nanotechnology Center has two Aligners, both are rigged for four inch wafers.
Safety Highlights:
1. If you haven't used the tool in a long time and have doubts about its operation, either ask a superuser to reacquaint you with it or work together with another knowledgeable user.
2. If you have any problems contact:
- Dan Haskell, 1125 Kemper Hall
- Bob Prohaska, 1125 Kemper Hall
- Lab Manager
Pre-operational Checklist
Before you enter the cleanroom:
1. Have you arranged training from the superuser or lab manager on the Aligner?
2. See the manual "Introduction to the Northern California Nanotechnology Center" for general reminders about clean room practice. If you have questions about where things are stored or hung, ask the lab manager or a superuser.
3. Do you have your tweezers?
First setup after entering the cleanroom:
1. Mark your use of this tool in the Aligner log book. Note any problems encountered by recent users.
2. Check to see if the mercury lamp is on. During the quarter it's best if the lamp remains on all the time. Turn on the mercury lamp thirty minutes to two hours before using the Aligner.
- a. Check to see that the Main Panel POWER is
OFF. (Fig. 4:
Main Panel). See Fig. 2: Top View to locate the
Main Panel.
- b. Turn ON the POWER switch (Fig. 6:
Mercury Lamp Power Supply). The mercury lamp power supply
is on the floor by the right front table leg.
- c. Press the START switch (Fig. 6).
- d. DO NOT change the CURRENT ADJUSTMENT setting (Fig. 6). If you think it needs adjustment, talk with the lab technician.
Setup after the mercury lamp has warmed up:
1. Turn on the house vacuum pump.
2. Check the pressure gauges (Fig. 2) to see if the nitrogen is on. If not, turn on the nitrogen. Generally, the nitrogen should always be on.
3. Turn ON the Aligner POWER (Fig. 4).
4. Turn ON the ILLUMINATOR. (Fig. 4).
5. Set the Main Panel switches (Fig. 4) as follows:
CONTACT/PROXIMITY: CONTACT
CONTACT: SOFT
WAFER FEED: AUTO.
6. If the MANU lamp is not lit, press the MANU button (Fig. 4).
7. Set the LIGHT INTEGRA thumb wheel (Fig.4) between 45- 60. Exposure time for these settings is roughly 4.5-6 seconds. You may need to experiment to find the optimum setting for your process.
8. Adjust the eyepieces:
- a. Twist left and right FOCUS ADJUSTMENT knobs
(Figs. 1-3) to
obtain the desired field illumination.
- b. Set lens separation for correct binocular
viewing:
- i. Read the eyepiece separation value from
between the eyepiece lenses.
- ii. Set the dial on the barrel of each eyepiece to match the eyepiece separation value.
- i. Read the eyepiece separation value from
between the eyepiece lenses.
Operating Instructions
1. Load and align the photomask.
- a. Unlock the saddle by pulling the Saddle
Release Lever
(Fig. 2). Swing the saddle to the right while holding
the saddle handle (see Figs. 1; and 3: Left
View).
- b. Fit the photomask, chrome or emulsion side
down, onto the photomask holder. Rest it against the three
guide pins (two on the back, one on the left).
- c. Press the MASK LOAD switch on the main panel (Fig. 4).
- d. Turn the photomask with the Photomask
Rotational
Adjustment knob (Fig. 1) so that the front edge of
the mask is roughly parallel to the front of the
Aligner.
- e. Swing the saddle back to its original
position.
- f. Use the Objective Spacing knobs (Figs. 1-3) to
find vertical scribe lines, i.e., the edge of the die. The
Objective Spacing knobs move the left and right
objective lenses horizontally.
- g. Turn the Photomask Rotational Adjustment
knob until the same patterns are seen in both fields of view.
Horizontal mask patterns should be parallel to a
line through the centers of both the right and left
objective lenses.
- h. Use the Scanning Grip (Fig. 3) to find the alignment structures (usually crosses). You will use these structures later to align wafers to the mask.
2. Load the wafer. For manual loading, see Subroutine 2: Manual Wafer Load below. For automatic loading, follow these steps:
- a. Place the wafer carriers with the H-bar side
down (see
Fig. 7: Wafer Carrier) on the Wafer Supply and
Take-up Stages (Fig. 2). Place unexposed wafers,
photoresist side up, in the supply carrier you placed
on the Supply Stage. The take-up carrier should be
empty.
- b. Press the START switch (Fig. 5: Front Panel). See (Fig. 2) to locate the Front Panel. The wafer moves to the pre-alignment station where it is automatically pre-aligned. Next, an auto hand swings the wafer from the pre-alignment station to the wafer chuck and sets the required alignment gap. Then the ALIGN lamp flashes (Fig. 5).
3. Make a coarse wafer alignment.
- a. Again unlock the saddle by pulling the
Saddle Release
Lever forward (Fig. 2). Swing the saddle to the
right while holding the saddle handle.
- b. With the naked eye make a rough alignment
with the
XY alignment grip (Fig. 2). Press in the black
button just above the grip to do a coarse alignment.
Release the button for a fine alignment.
On the front panel, press the top button on the Fine/Coarse Wafer Angle Control Selector (Fig. 5). This button determines the speed of rotational alignment.
- c. Now make a rough aligment with the Wafer
Angle
Control (Fig. 5). On the front panel, press the top
button on the Fine/Coarse Wafer Angle Control
Selector (Fig. 5). This button determines the speed of
rotational alignment.
- d. Return the saddle to its original position.
4. Align the wafer to the mask.
- a. If this is the first mask level for a
batch of wafers, see
Subroutine 1: First Mask Exposure below.
- b. If necessary, focus the left view field with
the Focus
Adjust Knob for both fields of view (Fig. 2). Then
focus the right view field with the right Focus
Adjust knob (Figs. 2 & 3).
- c. If the mask alignment structures are not in
both the left and right fields of view, repeat steps 1 (f) - (h)
above.
- d. Find an alignment structure on the wafer in
the left field of view using the XY alignment grip (Figs. 1 & 2). Make a
coarse alignment.
- e. Find a second alignment structure in the
right field of
view using the Wafer Angle Control (Fig. 5).
Make a coarse alignment.
- f. Align the wafer alignment structure in the
left view field to the mask alignment structure using the
XY alignment grip. Make first a coarse alignment
and then a fine one. Release the button on the grip
to do fine alignments.
- g. Align the wafer alignment structure in the
right field of view to the mask alignment structure
using the Wafer Angle Control (Fig. 5). Make
first a coarse alignment and then a fine one. Press
the appropriate button on the Fine/Coarse Wafer
Angle Control Selector to switch from coarse to fine
alignment (Fig. 5). This rotation will move the
wafer alignment structure in both the left and right
fields of vision. Therefore:
- h. Go back and repeat steps 4 (f) - (g) making
fine alignments until the alignment structures in both
fields of vision are aligned.
- i. Press the ALIGN switch (Fig. 5). The EXPO lamp will
start flashing. (Fig. 5). The wafer now rises and
makes contact with the mask. DO NOT move either
the wafer or the mask after they have made
contact.
- j. Check to see if the wafer and the mask are still correctly aligned. If the wafer has lost its proper alignment, press the ALIGN switch and repeat steps
4 (d) - (i).
5. Expose wafer.
- a. Attached to the frame of the scanning
grip (Fig. 3) is a
small horizontal grid and vertical pointer. Use the
scanning grip to move the pointer to the
approximate center of the grid. This centers the
exposure lens over the wafer.
- b. Press the EXPO switch (Fig. 5). The EXPO lamp lights, the ALIGN lamp goes out, and the wafer is exposed. After the wafer is exposed it is automatically ejected to the Take-up Stage and the next wafer is placed on the chuck ready for alignment.
Subroutine 1: First Mask Exposure
Because the first mask layer does not need coordination with previously aligned layers, one is free to expose it without alignment. Indeed, the Canon Mask Aligner can automatically expose an unpatterned batch of wafers without user assistance.
1. For the first wafer of a batch of unpatterned wafers, follow the operating procedure through step 3 (d).
2. Press the ALIGN switch.
3. Press the EXPO switch on the front panel (Fig. 5) and then immediately press the 1ST MASK switch on the main panel (Fig. 4). The Aligner will automatically pre-align, gap set, expose, and eject each succeeding wafer. Until the batch is finished, DO NOT touch the Wafer XY Alignment Adjustment knob (Figs. 1 & 2), the Wafer Angle Control (Fig. 5), or the Photomask Rotational Adjustment knob (Fig. 1).
Subroutine 2: Manual Wafer Load
Normally one loads the wafer so that the primary flat faces the rear of the aligner. By using manual load the flat may be oriented in any direction you desire. For a manual load, substitute the following procedure for the Operating Instructions step 2 above.
1. Set the Wafer Feed switch on the main panel to MANU (Fig. 4).
2. Using tweezers, place the wafer in the pre-alignment station oriented as you wish. On your left, find the pre- alignment station. It is under the auto hand at the end of the tracks leading from the Wafer Supply Stage. Allow for the rotation of the auto hand. The auto hand will rotate the wafer 130° clockwise when it loads the wafer.
Log Off
1. Leave the mercury lamp on if you plan to use this tool again within a week. In general, leave the lamp on throughout each quarter. This saves wear and tear on the bulb.
2. Remove your aligned wafers.
3. You're finished!
Troubleshooting
1. Check the logbook for any problems on prior runs.
2. If the auto arm fails to pick up the wafer from the pre-alignment stage, check whether you have turned on the vacuum pump. If the vacuum is on, check your wafer to see if it is uneven. If the wafer is not problematic, the suction cups on the auto arm may be the problem. Contact the lab technician for repairs.
3. For other problems, talk with the lab technician or check the manufacturer's service manual, section II. Find this manual in room 2136 Kemper Hall, filed under "Canon."